Cautions
Introduction
CAUTION:
Inaccurate readings may be caused by incorrect sensor
application or use; significant levels of dysfunctional
hemoglobins (i.e. carbohemoglobins or methemoglobin); or
intra-vascular dyes such as indocyanine green or methylene
blue; exposure to excessive illumination, such as surgical
lamps (especially ones with a Xenon light source), bilirubin
lamps, fluorescent lights, infrared heating lamps, or direct
sunlight; excessive patient movement; venous pulsations;
electro-surgical interference; and placement of a sensor on
an extremity that has a blood pressure cuff, arterial
catheter, or intra-vascular line.
CAUTION:
Route cables neatly. Ensure cables, hoses, and wires are
kept away from patient’s neck to avoid strangulation. Keep
floors and walkways free of cables to reduce risk to
hospital personnel, patients, and visitors. If the sensor or
patient cable is damaged in any way, discontinue use
immediately.
CAUTION:
When cleaning sensors, do not use excessive amounts of
liquid. Wipe the sensor surface with a soft cloth, dampened
with the cleaning solution. To prevent damage, do not soak
or immerse the sensor in any liquid solution. DO NOT
ATTEMPT TO STERILIZE.
CAUTION:
Recharge the Lithium ion battery while in the unit at room
temperature. If using the Accutorr V in a hot environment,
the Lithium ion battery may not charge when the unit is
connected to the AC mains.
CAUTION:
Remove the battery if the Accutorr V is not used for an
extended period of time.
CAUTION:
The Communications Connectors on the Accutorr V are only
for use with IEC 60601-1-1 compliant equipment.
CAUTION:
Never place fluids on top of this monitor. If fluid spills on the
unit, wipe clean immediately and refer the unit to qualified
service personnel.
CAUTION:
Before disassembling the monitor, eliminate static charges.
When disassembling the parts labeled with static-sensitive
symbols, wear electrostatic discharge protection such as
antistatic wristband or gloves. Follow the correct sequence
to disassemble the monitor. Otherwise, the monitor may be
damaged permanently. Disconnect all the cables before
disassembling any parts. Take care not to damage any
cables or connectors.
CAUTION:
Properly connect the cables or wires when reassembling the
monitor to avoid short circuit. When assembling the
monitor, select proper screws. If the wrong size screw is
tightened by force, the monitor may be damaged and the
screw or the part may not function as expected.
CAUTION:
All tests should be performed by qualified personnel only.
CAUTION:
Disconnect the monitor from the patient and make sure that
important data is saved before upgrading the monitor.
CAUTION:
Do not shut down or power off the equipment when
upgrading the bootstrap program. Otherwise, it may cause
the equipment to break down.