Disassembly
Repair and Disassembly
4 - 16
4.3.10
Disassembling the Temperature Module
1.
Remove the two (2) M3×6 cross-head screws (as shown in Figure 4-26).
2.
Take the cover plate off.
FIGURE 4-26
Remove Two (2) M3x6 cross-head screws
3.
Remove the three (3) M3×6 cross-head screws (as shown in Figure 4-27).
4.
Remove the Temperature module board.
5.
Remove the mylar sleeve.
FIGURE 4-27
Remove three (3) M3x6 cross-head screws.
M3x6 cross-head screw
M3x6 cross-head screw
M3x6 cross-head screw
M3x6 cross-head screw
M3x6 cross-head screw
Summary of Contents for Rosebud Accutorr V