Specifications
SpO
2
Measurement
2.4.2
Masimo MS-2013 SpO
2
Performance
This section details the performance measurements of the Masimo MS-2013, which is
designed in accordance with the requirements of the interface of the Masimo MS-2013
board. The Accutorr V interface complies with the requirements of the Masimo MS-2013
board communication protocol.
2.4.2.1
SpO
2
Compatible Sensors
:
See Section 6.6 for a list of compatible sensors.
Measurement range:
1 – 100%
Resolution:
1%
Accuracy no motion conditions:
Accuracy During motion conditions:
SpO
2
response time:
Under the condition that the PR is 75 BPM and the averaging is 8 seconds, the maximum
response time for the SpO
2
value to increase from 60% to 95% is 20 seconds.
Low Perfusion Performance:
PATIENT TYPE
SPO
2
RANGE: 70% – 100%
SPO
2
RANGE: 1% – 69%
Adult
±2 digits
Undefined
Pediatric
±2 digits
Undefined
Neonate
±3 digits
Undefined
PATIENT TYPE
SPO
2
RANGE: 70% – 100%
SPO
2
RANGE: 1% – 69%
Adult
±3%
Undefined
Pediatric
±3%
Undefined
Neonate
±3%
Undefined
LOW PERFUSION CONDITION
SPO
2
ACCURACY
PR ACCURACY
PULSE AMPLITUDE
% LIGHT
TRANSMISSIBILITY
> 0.02%
> 5%
± 2 digits
± 3 digits