6 - 13
Test and Calibration
SpO
2
Test
6.6
SpO
2
Test
NOTE:
A functional tester cannot be used to assess the accuracy of
a pulse oximeter probe or a pulse oximeter monitor.
Required Tool: SpO
2
simulator.
• For the monitor equipped with DPM SpO
2
module, BIO-TEK Index-2 SpO
2
simulator is
recommended.
• For the monitor equipped with Nellcor SpO
2
module, an SRC-MAX SpO
2
simulator is
recommended.
• For the monitor equipped with Masimo SpO
2
module, a BIO-TEK Index-2 SpO
2
simulator is recommended.
6.6.1
SpO
2
Test Under Normal Conditions
1.
Connect the SpO
2
simulator to the SpO
2
sensor.
2.
Select the model and the manufacturer of the SpO
2
module under test, and then
configure the SpO
2
simulator as follows: SpO
2
96%; PR 80 bpm.
3.
The displayed SpO
2
and PR values should be within the ranges listed below.
6.6.2
SpO
2
Test in Motion Mode
1.
Connect the SpO
2
simulator to the SpO
2
sensor.
2.
Select the model and the manufacturer of the SpO
2
module under test; take
measurement in the motion mode preset by the SpO
2
simulator.
3.
The displayed SpO
2
and PR values should be within the ranges listed below.
MANUFACTURER
SPO
2
SENSOR
SPO
2
PR (BPM)
DPM
Compatible Sensors
512E, 512G, 512F, 512H, 518B, 520A, 520P,
520I
96%±2%
80±3
520N
96%±3%
Masimo
Compatible Sensors
LNCS-NeoPt-L, LNCS Neo-L, LNOP DCI, LNOP
DCIP, LNOP TCI, LNOP YI-Multisite, LNOP DCSC,
LNOP Adt, LNOP Pdt, LNOP II Inf-L, LNOP II Neo-
Neonatal L, LNOP NeoPt Preterm Neonatal Y,
LNOP II NeoPt Preterm Neonatal Y
96%±3% (without
motion)
96%±3%
(with motion)
80±3 bpm (without
motion)
80±5 bpm
(with motion)
LNCS Inf-L, LNCS-Pdtx, LNCS-Adtx, LNCS DC-
I,LNCS DC-I Pt, TC-I
96%±2% (without
motion)
96%±3%
(with motion)
Nellcor
Compatible Sensors
MAX-A, MAX-P, MAX-I,
96%±2%
80±3 bpm
DS-100A, OXI-A/N (
Adult
), OXI-P/I, MAX-N
96%±3%
OXI-A/N (Neonate)
96%±4%
Summary of Contents for Rosebud Accutorr V