Testing the Optional Temperature Module
Test and Calibration
6 - 16
11.
Press
or
to change the selection to
°
C.
12.
Once
°
C is highlighted, press
to set it.
13.
Once the choices are set, press
to exit the
TEMP SETUP
dialog.
14.
Remove the Temperature probe from the probe sheath, insert a probe cover, and place
the probe into the oil tank.
15.
Wait till the Temp value displayed on the monitor stabilizes. Verify that the displayed
value is 37±0.1
°
C.
Contact Technical Support if the temperature test fails.