2 - 15
Specifications
Standby Mode
2.9
Standby Mode
• To enter the standby mode, the monitor must be disconnected from the patient and
receive no physiological signals from a patient with no low-battery alarm.
• To enter the standby mode, press
and confirm the operation in the pop-up dialog.
• After the monitor enters the standby mode:
• The LCD and the LEDs no longer display data.
• The SYS. LEDs Display flashes
— — —
.
• The current monitoring configuration and history data are saved.
• All current alarms are cleared.
• The NURSE CALL function is disabled. The NURSE CALL function is re-enabled when
the monitor resumes in normal mode.
• The connection with the central monitoring system stays the same as before entering
the standby mode.
• To resume normal mode:
1.
Press any key on the front panel.
2.
Confirm the resume operation at the confirm dialog. After confirming the resume
operation, the monitor exits the standby mode and enters the normal mode. If the
operation is not confirmed within 30 seconds, the system automatically selects Cancel,
and the monitor stays in the standby mode.
• In the following conditions, the monitor exits the standby mode and enters the normal
mode:
• The monitor receives SpO
2
physiological signals for 5 seconds.
• The predictive temperature sensor is removed from its sheath.
• The monitor is powered by battery, and a low-battery alarm occurs.
• Once the monitor enters the normal mode:
• The alarm function is re-enabled.
• The history data, patient information, and other settings from before the monitor
entered the standby mode are restored.
• Data communication is restored.
• The LCD displays.
• The LEDs display parameter data.
• Saving trend data starts.