Accutorr V
0070-10-0702
5 - 15
Par
ts
Main Brac
ke
t Assem
b
ly
NOTE:
**After replacing the Main Board, a System Reconfiguration should be performed by following this
procedure:
While the monitor is powered on, press and hold the MENU key for two seconds. While the MENU key is
still depressed, press and hold the power ON/OFF key. Verify that the message “Clear system config and
restarting” appears. Release the MENU and ON/OFF keys. The monitor will shutdown and restart. Then the
message “Checking system config...” appears. After the monitor is successfully configured, the message
“System configure succeed. Restarting...” is displayed and the monitor shuts down and restarts
automatically.