USER MANUAL 

Vital Signs Monitor 

Summary of Contents for Rosebud Accutorr V

Page 1: ...USER MANUAL Vital Signs Monitor...

Page 2: ...or a registered trademark of Shenzhen Mindray Bio Medical Electronics Co Ltd Nellcor is a U S registered trademark of Tyco International Ltd SmarTemp is a trademark of Shenzhen Mindray Bio Medical El...

Page 3: ...IBP Module 1 9 Optional Temperature Module 1 9 Recorder 1 10 Bar Code Scanner 1 10 Specifications 2 1 Introduction 2 2 NIBP Measurement 2 2 Blood Pressure 2 2 Static Accuracy 2 2 Maximum Cuff Pressure...

Page 4: ...ion 2 14 Multiple Level Volume 2 14 Real time Clock 2 14 Standby Mode 2 15 Alarm Information 2 16 Basics 2 16 Alarm Notification 2 16 Audio Alarm Pause Audio Alarm Off and Audio Alarm Silencing 2 16 C...

Page 5: ...rature Module 4 16 Disassembling SpO2 Modules 4 17 Disassembling a DPM SpO2 module 4 17 Disassembling a Masimo SpO2 module 4 18 Disassembling a Nellcor SpO2 module 4 19 Disassembling the NIBP Module 4...

Page 6: ...Nellcor SpO2 6 14 Masimo SpO2 6 14 Testing the Optional Temperature Module 6 15 Nurse Call Performance Test 6 17 Bar Code Scanner Test 6 19 Electrical Safety Tests 6 21 Enclosure Leakage Current Test...

Page 7: ...cludes NIBP Nellcor SpO2 a Trend Display and Recorder Accutorr V with Nellcor Pulse Oximetry and SmarTemp includes NIBP Nellcor SpO2 SmarTemp a Trend Display and Recorder Accutorr V with Masimo SET Pu...

Page 8: ...6 226 539 6 411 833 6 463 310 6 591 123 6 708 049 7 016 715 7 039 538 7 120 479 7 120 480 7 142 142 7 162 288 7 190 985 7 194 293 7 209 774 7 400 919 and 7 212 847 Possession or purchase of this devic...

Page 9: ...on the unit WARNING Possible explosion hazard Do not operate machine near flammable anesthetic agents or other flammable substances Do not use flammable anesthetic agents i e ether or cyclopropane WAR...

Page 10: ...cted to medical equipment within patient environments must meet the requirements of the IEC 60601 1 1 medical electrical systems standard and should be put under electrical safety tests at the frequen...

Page 11: ...ng the Accutorr V in a hot environment the Lithium ion battery may not charge when the unit is connected to the AC mains CAUTION Remove the battery if the Accutorr V is not used for an extended period...

Page 12: ...perating controls can easily be accessed NOTE The instructions in this manual are based on the maximum configuration NOTE The optional Temperature module kit must be installed only by trained personne...

Page 13: ...rnal electric power source Degree of protection against electric shock Monitor Type B applied part NIBP Type BF defibrillation protected applied part SpO2 Type BF protected applied part Temp Type BF p...

Page 14: ...ral pulses taken only during a measurement cycle The rate measured by the Accutorr V may differ from the rate of an ECG monitor This is because the ECG is an electrical signal that may not always resu...

Page 15: ...in resolving shipping problems NOTE The Accutorr V should only be shipped in its original packing materials to avoid shipping damage Symbols and Descriptions SYMBOL DESCRIPTION SYMBOL DESCRIPTION Atte...

Page 16: ...ctor NIBP interval key SP1 RS 232 connector Serial Port 1 Print key front panel CS1 Network connector Print key recorder Alarm Disabled indicator on LCD display Alarm Silenced indicator on LCD display...

Page 17: ...Accutorr V Service Manual 0070 10 0702 1 1 1 0 Theory of Operation Introduction 1 2 Controls and Indicators 1 2 System Overview 1 2 Hardware Structure 1 4...

Page 18: ...icators refer to the Accutorr V Operating Instructions Part Number 0070 10 0699 02 1 3 System Overview FIGURE 1 1 shows the relationship of the Accutorr V monitoring system s mechanical hardware and s...

Page 19: ...ctions J10 J3 J2 J1 J4 J6 J2 P1 P2 J2 J1 J1 P5 J6 J2 J2 J7 J8 J9 J2 J3 J4 J5 J11 J12 J13 P3 P4 P2 J4 J5 POWER RECORDER SpO2 TEMP NIBP ISOLATION POWER BOARD MAIN BOARD PARAMETER RECEPTACLE KEY DISPLAY...

Page 20: ...ain board processes measurements and the status of all modules and displays them through the key and display board FIGURE 1 3 The Accutorr V s Hardware Structure NOTE The SpO2 isolation power board do...

Page 21: ...ace The main board controls the alarm indicator using FPGA The main board controls the speaker to give audible alarms key tones and Pitch Tone The main board provides the nurse call connection network...

Page 22: ...verters and charging circuit The DC DC converters convert 16 8V DC to 12V 5V and 3 3V DC The charging circuit charges the lithium battery When the Accutorr V is not connected to the AC mains the batte...

Page 23: ...te LCD display NOTE The monitor with color LCD does not have Contrast setting item in the common setup menu The 7 segment digit units display parameter data The AC indicator is driven by the ADV outpu...

Page 24: ...s transmitting through the finger or toe and processes the collected signals to create the measured result The SpO2 module controls the LED drive circuit and the amplifying circuit gain corresponding...

Page 25: ...onal Temperature Module Principle Diagram The Temperature Module uses a thermistor as a sensor for measuring temperature The resistance of a given thermistor is nonlinearly relative to the temperature...

Page 26: ...lowing table The monitor s serial port is defined in the following table Basic settings of the bar code scanner are listed in the following table PIN DEFINITION 2 Barcode_RX 3 Barcode_TX 5 GND 9 VCC5V...

Page 27: ...0070 10 0702 1 11 Theory of Operation Hardware Structure To change parameter settings scan the following bar codes in sequence 1 Set parameter defaults 2 Set beeper volume 3 Set trigger mode 4 Set sca...

Page 28: ...Hardware Structure Theory of Operation 1 12 0070 10 0702 Accutorr V Service Manual 5 Disable parameter scanning...

Page 29: ...ion 2 2 NIBP Measurement 2 2 NIBP Measurement Cycle 2 4 SpO2 Measurement 2 6 Temperature 2 10 Display Area Indicator and Controller 2 10 Audio Indicator 2 14 Real time Clock 2 14 Standby Mode 2 15 Ala...

Page 30: ...isons with auscultatory values measured using the fifth Korotkoff sound within ANSI AAMI SP10 standards for accuracy 2 2 1 Blood Pressure 2 2 2 Static Accuracy 2 2 3 Maximum Cuff Pressure Normal Use O...

Page 31: ...ive Blood Pressure Cuff venting mechanism in adult mode vents a volume of 500 cc reduced from a pressure of 260 mmHg to a pressure of 15 mmHg in a maximum of 10 seconds The Non Invasive Blood Pressure...

Page 32: ...ssure of the last measurement If the first measurement is unsuccessful the subsequent inflation pressures change to 50 mmHg higher than the previous systolic pressure in adult or pediatric mode and 40...

Page 33: ...the systolic value of the previous reading After the first unsuccessful measurement the subsequent inflation value will be the previous systolic pressure reading plus P2 2 3 3 Pulse Rate Accuracy 3 BP...

Page 34: ...e measured with a CO oximeter Pulse oximeter measurements are statistically distributed and only about two thirds of the measurements can be expected to fall within the specified accuracy compared to...

Page 35: ...e sensors Measurement range 1 100 Resolution 1 Accuracy no motion conditions Accuracy During motion conditions SpO2 response time Under the condition that the PR is 75 BPM and the averaging is 8 secon...

Page 36: ...repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3cm in induced hypoxia studies in the range of 70 to 100 SpO2 against a laboratory co oximeter and ECG monitor This variation equals plus or...

Page 37: ...on neonatal subjects as recommended the specified accuracy range increases by 1 digit to account for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal blood 2 Neonatal a...

Page 38: ...32 C 89 6 F not including 32 C 89 6 F 0 2 C 0 4 F 60 s 32 C 89 6 F 44 C 111 2 F including 32 C 89 6 F 0 1 C 0 2 F predict 35 C 43 C 95 F 109 4 F Not defined typically 10 s 12 s from the moment when te...

Page 39: ...screen menu fonts and LCD prompting characters can be changed through a software upgrade The default start logo is Accutorr V The format of the date can be set in the maintenance mode Chinese style y...

Page 40: ...Press to start an NIBP measurement Deflate During an NIBP measurement stops the measurement Also suspends interval mode until the next time NIBP Start is pressed Print Start recording trend data real...

Page 41: ...as used on Trio to the monitor configured with the DPM SpO2 board Apply the Masimo connector to the monitor configured with the MS 2013 SpO2 board The monitor configured with the Masimo SpO2 board mus...

Page 42: ...ted according to the SpO2 value and the frequency increases as the SpO2 value increases The monitor can perform automatic 22 level modulation of the pulse tone s pitch 2 7 2 Multiple Level Volume Ther...

Page 43: ...e standby mode To resume normal mode 1 Press any key on the front panel 2 Confirm the resume operation at the confirm dialog After confirming the resume operation the monitor exits the standby mode an...

Page 44: ...notifies the user of the alarm with the highest level through the speaker and alarm indicator 3 If the hospital Nurse Call system is available and connected to the monitor the user can be notified of...

Page 45: ...Accutorr V Service Manual 0070 10 0702 3 1 3 0 Troubleshooting Introduction 3 2 Part Replacement 3 2 Monitor Status Check 3 2 Technical Alarm Check 3 3 Troubleshooting Guide 3 4...

Page 46: ...t and retest If the trouble remains exchange the replacement PCB with the original suspicious PCB and continue troubleshooting as directed in this chapter 3 3 Monitor Status Check Some troubleshooting...

Page 47: ...TIME or MODULE VERSION 7 Once the selection is highlighted press to display the data 8 Press to exit to the MAINTENANCE dialog 3 4 Technical Alarm Check If a technical alarm message displays on the L...

Page 48: ...and LED board Power module defective Replace the power module Main board Defective Replace the main board SYMPTOMS POSSIBLE CAUSE CORRECTIVE ACTIONS The display is black or blank but the monitor stil...

Page 49: ...nce MAINTENANCE is highlighted press to display the MAINTENANCE dialog as shown in FIGURE 3 2 4 Press or to select USER MAINTENANCE 5 Once USER MAINTENANCE is highlighted press to display the USER MAI...

Page 50: ...ad and the paper roller do not contain foreign matter Clean the thermal print head with an appropriate detergent Recorder failure Replace the recorder SYMPTOMS POSSIBLE CAUSE CORRECTIVE ACTION No nurs...

Page 51: ...one fails Cable defective or poorly connected Confirm that the cable between battery interface board and the power module is properly connected Confirm that the connecting cables and connectors are n...

Page 52: ...FIGURE 3 1 2 Press or to highlight MAINTENANCE 3 Once MAINTENANCE is highlighted press to display the MAINTENANCE dialog as shown in FIGURE 3 2 4 Press or to select USER MAINTENANCE 5 Once USER MAINT...

Page 53: ...d Replace the NIBP module Air tubing occluded Replace the air tubing SYMPTOMS POSSIBLE CAUSE CORRECTIVE ACTIONS Unable to measure SpO2 SpO2 sensor failure Replace the SpO2 sensor Cables defective Conf...

Page 54: ...Troubleshooting Guide Troubleshooting 3 10 0070 10 0702 Accutorr V Service Manual This page intentionally left blank...

Page 55: ...sassembling the Front Housing Assembly 4 6 Removing the Main Rack Assembly 4 9 Removing NIBP Module 4 11 Removing the Main CPU Board 4 12 Removing the Power Board 4 13 Removing the Battery Connector A...

Page 56: ...efore disassembling the monitor eliminate static charges When disassembling the parts labeled with static sensitive symbols wear electrostatic discharge protection such as antistatic wristband or glov...

Page 57: ...l Temperature module proceed to step 2 If the monitor is configured with an Optional Temperature module a Remove the two 2 M3 6 cross head screws as shown in Figure 4 1 FIGURE 4 1 Remove two 2 M3x6 cr...

Page 58: ...r 4 M3 8 cross head screws as shown in Figure 4 3 FIGURE 4 3 Remove four 4 M3x8 cross head screws 3 Release the clips as shown in Figure 4 4 on the front and rear housings FIGURE 4 4 Release the clips...

Page 59: ...he front and the rear housings 5 Separate the front and the rear housings Figure 4 5 FIGURE 4 5 Separate the front and rear housings NOTE Exercise care when separating the front and the rear housings...

Page 60: ...the key board Black and White LCD Front Housing Assembly Color LCD Front Housing Assembly FIGURE 4 6 Remove six 6 M3x6 cross head screws M3x6 cross head screw M3x6 cross head screw M3x6 cross head scr...

Page 61: ...ep FIGURE 4 7 Remove four 4 M3x8 cross head screws 3 Disconnect the flexible cables on the button board 4 Remove the LCD 5 Remove the two 2 M3 6 cross head screws as shown in FIGURE 4 8 and take out t...

Page 62: ...w 7 Remove the two 2 screws for the SpO2 signal cable as shown in Figure 4 10 8 Take out the SpO2 signal cable FIGURE 4 10 Remove two 2 screws 9 For DPM SpO2 only unscrew the plastic nut for the DPM S...

Page 63: ...s head screws as shown in Figure 4 12 2 Take out the recorder by pushing the clips inward FIGURE 4 12 Remove two 2 M3x6 cross head screws 3 Unplug the two 2 cable connectors as shown in Figure 4 13 fr...

Page 64: ...an optional Temperature module remove the Temperature module board FIGURE 4 14 Remove optional temperature module board 6 Remove the six 6 M3 6 cross head screws as shown in Figure 4 15 7 Take out the...

Page 65: ...IBP module from the CPU Board as shown in Figure 4 16 FIGURE 4 16 Unplug the NIBP module 2 Remove the three 3 M3 6 cross head screws as shown in Figure 4 17 3 Take out the NIBP rack FIGURE 4 17 Remove...

Page 66: ...e connector and cut one 1 tie wrap 3 Unplug the speaker cable J10 from the Main board see FIGURE 4 19 4 Unplug the recorder cable J8 from the Main board see FIGURE 4 19 5 Take out the main board by re...

Page 67: ...Unplug the remaining cable connectors fan AC mains connector and Main CPU board from the Power board FIGURE 4 20 Remove three 3 M3x6 cross head screws 4 3 7 Removing the Battery Connector Assembly Rem...

Page 68: ...the main rack 2 Remove the fan cushion as shown in Figure 4 22 FIGURE 4 22 Remove the Fan Cushion 3 Remove the three 3 M3 25 cross head screws as shown in Figure 4 23 4 Remove the fan 5 Remove the sp...

Page 69: ...ining nut securing the grounding post as shown in Figure 4 24 2 Remove the grounding terminal FIGURE 4 24 Remove the retaining nut 3 Remove the two 2 M3 8 cross head screws as shown in Figure 4 25 4 R...

Page 70: ...igure 4 26 2 Take the cover plate off FIGURE 4 26 Remove Two 2 M3x6 cross head screws 3 Remove the three 3 M3 6 cross head screws as shown in Figure 4 27 4 Remove the Temperature module board 5 Remove...

Page 71: ...sembling SpO2 Modules 4 3 11 1 Disassembling a DPM SpO2 module 1 Remove the M3 6 cross head screw as shown in FIGURE 4 28 FIGURE 4 28 Remove one M3 X 6 cross head screw 2 Remove the SpO2 board as show...

Page 72: ...ower board by removing the three 3 M3 nuts as shown in Figure 4 30 2 Remove the three 3 M3 6 cross head screws FIGURE 4 30 Remove three 3 M3 nuts and three 3 M3x6 cross head screws 3 Remove the SpO2 b...

Page 73: ...ply 4 3 11 3 Disassembling a Nellcor SpO2 module 1 Remove the M3 6 cross head screw as shown in FIGURE 4 33 to separate the SpO2 module from its isolated power supply as shown in FIGURE 4 34 FIGURE 4...

Page 74: ...t the two tie wraps holding the NIBP pump as shown in FIGURE 4 35 4 Remove the pump FIGURE 4 35 Cut two 2 tie wraps 4 3 12 2 Removing the NIBP Dump Valve 1 Unplug the dump valve cable connector from t...

Page 75: ...assembly Disassembly 4 3 12 3 Removing the NIBP Bleed Valve 1 Unplug the bleed valve cable connector from the NIBP board 2 Disconnect the tubing 3 Remove the two 2 screws as shown in FIGURE 4 37 4 Rem...

Page 76: ...Disassembly Repair and Disassembly 4 22 0070 10 0702 Accutorr V Service Manual This page intentionally left blank...

Page 77: ...Main Assembly 5 2 Front Assembly Color LCD 5 3 Rear Housing Assembly 5 12 Main Bracket Assembly 5 13 NIBP Assembly Pump M6Q 100003 5 16 NIBP Assembly Pump 082 000056 00 5 19 Temperature Module Subass...

Page 78: ...nt Shell Assembly B W Front Shell Assembly Color 2 M6G 020015 Hose Silicone 1 6mmODX0 8mmID 10m per 3 M04 004012 Screw Crosshead M3 6 4 6101 30 46616 Main Bracket Assembly 5 801 6101 00011 00 Recorder...

Page 79: ...Assembly Service Kit Color LCD Includes all parts below 11 0010 20 12194 Machine Pneumatic Socket Subassembly 12 043 001894 00 Front Bezel Color LCD 13 6101 20 46722 Alarm Light Cover 6101 14 6101 20...

Page 80: ...FRU 801 6101 00023 00 Included in this FRU are FRU 801 6101 000211 00 Main board 009 002037 00 LCD signal cable 042 004637 00 Ferrite Fixer Plate 051 000875 00 Display and Key panel 049 000280 00 Colo...

Page 81: ...26 801 6101 00016 00 Masimo SpO2 Board 27 801 6101 00015 00 Masimo Power Isolation Board PCBA 28 801 0010 00019 00 Masimo SpO2 Flex Cable 29 6101 20 46720 51 Masimo SpO2 Interface Bracket 25 19 26 27...

Page 82: ...te Fixer Plate Isolation Sheet 38 042 004637 00 Ferrite Fixer Plate Color LCD FRU 801 6101 00023 00 contains these parts 39 0000 10 10996 Spring Strip Upgrading the Black and White LCD assembly to the...

Page 83: ...0 047 005564 00 047 005565 00 Front Overlay 6101 English Front Overlay 6101 Italy Front Overlay 6101 French Front Overlay 6101 German Front Overlay 6101 Spanish Front Overlay 6101 Portuguese Front Ove...

Page 84: ...3 00 Front Assembly Service Kit Black White Includes all parts below 11 0010 20 12194 Machine Pneumatic Socket Subassembly 44 6101 20 46710 Front Bezel 6101 13 6101 20 46722 Alarm Light Cover 6101 14...

Page 85: ...20 46719 51 Nellcor SpO2 Interface Bracket 25 6101 30 46621 Masimo SpO2 Module Package 26 801 6101 00016 00 Masimo SpO2 Board 25 19 26 27 28 29 30 23 22 21 20 24 19 ITEM PART NUMBER DESCRIPTION 27 80...

Page 86: ...6 45 M04 011002 Hex Locknut with Conical Washer M3 46 0000 10 10997 LCD Module FSTN 320x160 LED White Backlight 47 6101 30 46609 Display Keyboard PCBA 48 6101 20 46752 Ferrite Fixer Plate 49 6101 20 4...

Page 87: ...CRIPTION 3 M04 004012 Screw Crosshead M3 6 53 6101 20 46732 Grounding Washer 6101 54 047 002718 00 6101 20 46741 52 6101 20 46741 53 6101 20 46741 54 6101 20 46741 55 6101 20 46741 56 Front Overlay 61...

Page 88: ...r Service Kit Includes all parts below except 801 6101 00009 00 51 M04 021024 Large Washer 55 DA6H 20 22831 Foot Cushion 56 801 6101 00009 00 Battery Door 6101 and Battery Door Link 57 M04 000802 Flat...

Page 89: ...ing 62 ITEM PART NUMBER DESCRIPTION 69 7000 20 24365 Torsion Spring 70 6006 20 39486 Battery Bar Setscrew 71 6101 20 46739 Fan Gasket 6101 72 M04 051172 Screw Flat Head Phillips M3X25 73 6101 20 46731...

Page 90: ...D Signal Cable for B W LCD LCD Signal Cable for color LCD 93 M09A 20 62081 Communication Cable from Host Board to Isolation Board 94 6006 20 39382 SpO2 Module Cable 95 6006 20 39383 Keyboard Signal Ca...

Page 91: ...ey for two seconds While the MENU key is still depressed press and hold the power ON OFF key Verify that the message Clear system config and restarting appears Release the MENU and ON OFF keys The mon...

Page 92: ...Service Manual 5 6 NIBP Assembly Pump M6Q 100003 Accutorr V models with serial numbers lower than S N A7503017K9 are configured with the old NIBP pump 083 000056 00 FIGURE 5 13 NIBP Assembly Pump M6Q...

Page 93: ...019 00 NIBP PCBA Module 100 A21 000002 Tubing Silicone 1 8 X1 4 X100ft 2800546 100 101 S1 0378 02 0004 Filtered Orifice Restricto 102 S1 0008 10 0206 Tubing Silicone 1 16 X3 16 TYGON 3350 1foot 103 30...

Page 94: ...BP Assembly Pump 082 000056 00 for the old pump assem ITEM PART NUMBER DESCRIPTION 107 047 001582 00 NIBP Bracket Label 108 M04 051137 Screw M2 4 109 M90 000002 00 Insulation Washer F2 110 042 001417...

Page 95: ...ly Pump 082 000056 00 Parts 5 19 0070 10 0702 Accutorr V Service Manual 5 7 NIBP Assembly Pump 082 000056 00 FIGURE 5 16 NIBP Assembly Pump 082 000056 00 ITEM PART NUMBER DESCRIPTION 10 M04 002505 Scr...

Page 96: ...0 00019 00 NIBP PCBA Module 100 A21 000002 Tubing Silicone 1 8 X1 4 X100ft 2800546 100 101 S1 0378 02 0004 Filtered Orifice Restricto 102 S1 0008 10 0206 Tubing Silicone 1 16 X3 16 TYGON 3350 1foot 10...

Page 97: ...6 pg 5 16 NIBP Assembly Pump M6Q 100003 for the new rolling pump assembly ITEM PART NUMBER DESCRIPTION 116 6101 20 46737 NIBP Fram Overlay 6101 108 M04 051137 Screw M2 4 109 M90 000002 00 Insulation...

Page 98: ...erature Module Subassembly FIGURE 5 19 Temperature Module Subassembly ITEM PART NUMBER DESCRIPTION 10 M04 002505 Screw M3 6 119 6101 20 46759 Temp Module Water Ingress Blocker 120 801 6006 00044 00 Te...

Page 99: ...e Manual FIGURE 5 20 Temperature Module Subassembly ITEM PART NUMBER DESCRIPTION 122 6101 20 46717 Temp Probe Cover Holder 123 6101 20 46715 Temp Module External Cover 124 6101 20 46724 Temperature Pr...

Page 100: ...ssembly FIGURE 5 21 Temperature Power Module Subassembly ITEM PART NUMBER DESCRIPTION 10 M04 002505 Screw M3 6 126 M09A 20 62082 Communication Cable from Power Isolation Board to Temp Measurement Boar...

Page 101: ...ion 6 2 Visual Test 6 4 Power on Test 6 4 NIBP Calibration 6 5 NIBP Accuracy Test 6 9 SpO2 Test 6 13 Testing the Optional Temperature Module 6 15 Nurse Call Performance Test 6 17 Bar Code Scanner Test...

Page 102: ...el only NOTE Qualified personnel should acquaint themselves with the test tools and confirm that test tools and cables are applicable NOTE When fields or buttons in a dialog have a line through them t...

Page 103: ...bar code scan is incorrect Electrical safety tests Enclosure leakage current test 1 Following any repair or replacement of the power module 2 At least once every two years Earth leakage current test P...

Page 104: ...ble 6 3 Power on Test Verify that the monitor can power up correctly by following this procedure 1 Insert the lithium battery in the battery compartment and connect the monitor to the AC mains The AC...

Page 105: ...1 FIGURE 6 1 NIBP calibration configuration 2 If needed press 11 to display the Normal Screen 3 Press 12 to display the SYSTEM SETUP dialog 4 Press 16 or 19 to highlight MAINTENANCE 5 Once MAINTENANC...

Page 106: ...og 10 In the Enter Password dialog press 18 See FIGURE 6 4 11 Press 16 or 19 to set the first password digit to 3 12 Press 18 13 Press 16 or 19 to highlight the second password digit 14 Press 18 15 Pr...

Page 107: ...to highlight CANCEL then press 18 to cancel the calibration NOTE After the calibration vessel is pressurized the reference pressure field is highlighted as shown in FIGURE 6 6 FIGURE 6 6 NIBP CALIBRA...

Page 108: ...P CALIBRATION dialog with ACCEPT highlighted 26 Press the 16 or 19 to select ACCEPT 27 Press 18 to accept the calibration NOTE Use 16 or 19 to highlight CANCEL then press 18 to cancel the calibration...

Page 109: ...E 6 8 FIGURE 6 8 NIBP accuracy test configuration 2 If needed press to display the Normal Screen 3 Press to display the SYSTEM SETUP dialog 4 Press or to highlight MAINTENANCE 5 Once MAINTENANCE is hi...

Page 110: ...ch the value shown on the reference manometer within the accuracy for each test range as shown in the table 11 Use or to highlight STOP ACCURACY TEST then press to end the accuracy test If the accurac...

Page 111: ...guration 3 If needed press to display the Normal Screen 4 Press to display the SYSTEM SETUP dialog 5 Press or to highlight MAINTENANCE 6 Once MAINTENANCE is highlighted press to display the MAINTENANC...

Page 112: ...button changes to STOP LEAK TEST Use or to highlight STOP LEAK TEST then press to cancel the leak test 11 If the system leaks the message PNEUMATIC LEAK will be displayed in the technical alarm area p...

Page 113: ...should be within the ranges listed below 6 6 2 SpO2 Test in Motion Mode 1 Connect the SpO2 simulator to the SpO2 sensor 2 Select the model and the manufacturer of the SpO2 module under test take meas...

Page 114: ...Summary of Test Methods 6 6 3 1 DPM SpO2 Measurement validation The DPM SpO2 module accuracy has been validated in human studies against arterial blood sample reference measured with a CO oximeter Pul...

Page 115: ...UP dialog 3 Press or to highlight TEMP SETUP 4 Once TEMP SETUP is highlighted press to display the TEMP SETUP dialog as shown in FIGURE 6 14 FIGURE 6 14 Example TEMP SETUP dialog 5 Press or to highlig...

Page 116: ...ighted press to set it 13 Once the choices are set press to exit the TEMP SETUP dialog 14 Remove the Temperature probe from the probe sheath insert a probe cover and place the probe into the oil tank...

Page 117: ...g 8 Press or to highlight the SIGNAL DURATION pull down list 9 Once the SIGNAL DURATION pull down list is highlighted press to select it 10 Press or to highlight the PULSE 11 Once the PULSE is highlig...

Page 118: ...N pull down list is highlighted press to select it 23 Press or to highlight the CONTINUOUS 24 Once the CONTINUOUS is highlighted press to select it 25 In the NURSE CALL SETUP dialog set SIGNAL TYPE to...

Page 119: ...TENANCE 4 Once MAINTENANCE is highlighted press to display the MAINTENANCE dialog as shown in FIGURE 6 16 FIGURE 6 16 MAINTENANCE dialog 5 Press or to select USER MAINTENANCE 6 Once USER MAINTENANCE i...

Page 120: ...selection is highlighted press select it NOTE When the RS 232 connector is used for DIAP barcode power must be set to OFF 13 Aim the bar code scanner at the target bar code Adjust the field of view t...

Page 121: ...trical hazards to the patient operator or service personnel Electrical safety tests should be performed under normal ambient conditions of temperature humidity and atmospheric pressure The electrical...

Page 122: ...ly DC leakage reading is followed by DC Under normal status the patient leakage current should be no greater than 100 A AC 10 A DC Under single fault condition the leakage current should be no greater...

Page 123: ...The recorder should print correctly and printout should be clear 2 Open the recorder door and verify the monitor gives the proper message Close recorder door before proceeding to the next step 3 Set...

Page 124: ...rade package is the correct one To obtain the latest upgrade package contact Service The following is a list of upgrade programs Bootstrap program System program Bar code recognition configuration Mul...

Page 125: ...0070 00 0702 Rev 7 0 October 2015...