Making Measurements
Measuring Conversion loss
Normalization is the simplest type of calibration. The analyzer stores
normalized data into memory and divides subsequent measurements by the
stored data to remove unwanted frequency response errors. This calibration
is used for this measurement to remove the insertion loss error of the IF
filter. Changing the frequency span or number of measurement points will
invalidate a normalization calibration.
Perform the following steps to perform a normalization calibration:
1. Connect the equipment as shown in Figure 3-14, except replace the mixer
with a through cable.
2. Set the following frequency parameters:
Center
MHz
Span
This sets the analyzer frequency range to sweep over the
of the
IF filter (700 MHz).
Press
Normalize
This stores the filter response
into memory, and sets up a
normalized trace so that the filter response magnitude is removed from the
measurement.
4. Replace the through cable with the mixer as shown in Figure 3-14.
Press
Center
MHz to change the center frequency so that the
mixing product of the mixer is in the
of the IF filter.
3-38
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...