Calibrating far Increased Measurement
Accuracy
Choose an Appropriate Calibration Method
Normalization is the simplest type of calibration. The analyzer stores data
into memory and divides subsequent measurements by the stored data to
remove frequency response errors.
Follow these general steps when performing a normalization calibration:
1.
2.
3.
4.
Setup the analyzer for your measurement:
Select the type of measurement
Enter operating parameters other than the default
Connect your equipment as you would for an actual measurement, but
omit the DUT.
Notice that the top of the display on the analyzer now shows the type
of measurement followed by
to indicate that the displayed data is
actually the measurement divided by memory.
C A U T I O N
The normalization cal will be invalidated if any frequency settings are
changed after calibration. There is no correction interpolation with a
normalization cal .
6-10
C A U T I O N
If you decide to perform any other type of calibration, be sure to turn off
normalization
by pressing
. Note that the
indicator
disappears when normalization is turned off.
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...