I
Making Measurements
Measuring Transmission Response
1. To view the entire measurement trace on the display, press
2. To interpret the transmission measurement, refer to Figure 3-6 or
your analyzer’s display if you are making this measurement on your
instrument.
a. The values shown on the horizontal axis are the frequencies in MHz.
The values shown on the vertical axis are the power ratios in decibels
of the transmitted signal through the device divided by the
incident power. To display the result in logarithmic magnitude format
(designated by “Log Mag” at the top of the measurement screen), the
analyzer computes the measurement trace using the following formula:
Transmission
=
10
log
where
= the power transmitted through the device and
where
= the incident power.
b. A level of 0
would indicate a perfect through cable or device (no
loss or gain). Values greater than 0
indicate that the DUT has gain.
Values less than 0
indicate loss.
3. To quickly determine the filter’s minimum insertion loss, press
Marker
Search Mkr
Max.
3-22
4. Note the marker readout in Figure 3-6 provides the frequency and
amplitude of the minimum insertion loss point.
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...