Automating Measurements
Configuring Your Test System
Operation of an
Commands can be sent from an external computer or from
or both at
Analyzer Running
the same time, if certain precautions are observed. Things to consider:
Under Computer
Control
1. If both the analyzer and the computer send SCPI commands at the same
time, the analyzer may not Enish the
command before executing
the computer’s command, or vice-versa. The programmer must ensure
that SCPI commands executed by
do not overlap with SCPI
commands sent from an external controller, otherwise the system may
deadlock. Synchronization between the analyzer and the controller must
be ensured. See “Synchronizing the Analyzer and a Controller” in the
supplement to the
HP Instrument BASIC User’s Handbook
for additional
information.
2. Both
8) and the external controller (LAN or
7) share the same HP-IB status model (the same analyzer status bits go to
each). Be careful sending commands which affect status reporting, such as
STAT:PRES,
etc.
7-14
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...