I
Calibrating for Increased Measurement
Accuracy
Check the Calibration
Load Match
Load match is a transmission measurement error term. It is
a measure of the re-reflections contributed from the match
of the THRU standard and the receiver port (RF IN).
A typical value for this match term is 10
L o g
5.0
Ref -45.00
O f f
- 2 5
I
I
I
I
I
I
I
I
I
S t a r t 0 . 3 0 0 M H Z
1 3 0 0 . 0 0 0 M H Z
Figure 6-7. Typical load Match Error Term
6 - 3 3