Using Instrument Functions
Using Markers
To Use Marker Flatness
The marker flatness search function measures a user-defined segment of the
measurement trace and calculates the following:
frequency span
gain
slope
flatness
The analyzer calculates flatness by drawing a straight line between the
markers. A maximum vertical deviation from this line is computed for each
measurement point. Flatness is the magnitude difference of the maximum
and minimum calculated deviations from the straight line.
Limit testing may be performed on the flatness parameter. See “lb Use
Marker Limit Functions, later in this chapter for information.
1. On measurement channel 1 press
and set markers 1 and 2
to
the beginning and end of the trace segment that you want to
measure. (When using measurement channel 2, use markers 3 and 4 to
define the trace segment.)
Press
Functions.
Math
.
3. Figure 4-13 shows a
trace segment. Notice the marker readout in
the upper right corner of the display.
4-22
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...