Making Measurements
Measuring Reflection Response
Your analyzer can provide highly accurate measurements without performing
any additional user-calibrations if certain conditions are met. This example
describes how to perform a reflection one-port calibration. A one port
calibration uses known standards to correct for directivity, source match, and
frequency response errors in narrowband measurements.
To perform a reflection one-port calibration you will need one of the following
calibration kits depending on the nominal impedance of your analyzer:
HP 850323
for 50 type-N female connector calibrations
HP
for type-N female or type-N male 50 connector calibrations
HP 850363
for 75 type-N female connector calibrations
HP
for type-N female or type-N male 75 connector calibrations
HP
for 3.5 mm female or 3.5 mm male 50 connector
calibrations
HP
for Type-F 75 connector calibrations
N O T E
If you are going to be using calibration standards other than the default
type-N), you must
select the type of connector type by pressing (CAL)
Cal
Kit and then selecting the appropriate
3 - 2 5
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...