Installing the Analyzer
Step 4. Configure the Analyzer
HP-IB Connections
An HP-IB system may be connected in any configuration as long as the
following rules are observed:
l
The total number of devices is less than or equal to 15.
l
The total length of all the cables used is less than or
to 2 meters
times the number of devices connected together up to an absolute
maximum of 20 meters. For example, the maximum cable length is
4 meters if only 2 devices are involved. The length between adjacent
devices is not critical as long as the overall restriction is met.
See Figure
for different connection
S T A R
L I N E A R
NETWORK
NETWORK
ANALYZER
ANALYZER
Figure 1-5. HP-IB Connection Configurations
1-13