I
Using Instrument Functions
Using limit Testing
Delta Amplitude
This marker limit test allows you to set marker 1 as an amplitude reference
against which marker 2 is limit tested.
1.
2.
3.
4.
5.
This limit test requires that you Erst use marker 1 to determine the
reference amplitude: Press
2: and then use the front panel
knob or the
keys to place marker 1 at the desired place on the
measurement trace. The amplitude of marker 1 at this point becomes the
delta reference for this marker limit test.
Press
Limit
Limits
Use the front panel knob or the
keys to select Delta
in
the marker limit test table. Turn this limit function on by pressing the
OFF key. Note that the entry in the on/off column of the
table changes to
Press Edit Limit
Limit and enter the maximum
limit using the front panel keypad and terminating the entry with the
k e y .
Press Edit Limit
and enter the minimum limit
using the front panel keypad and terminating the entry with the
key.
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...