Calibrating for Increased Measurement
Accuracy
Choose an Appropriate Calibration Method
Conversion loss measurements typically utilize only a normalization
calibration. See
Perform a Normalization Calibration” earlier in this
chapter.
When in conversion loss measurement mode, the analyzer is using its internal
broadband detectors and pressing the (CAL) key calls up a menu for zeroing
the detectors. Zeroing the detectors helps compensate for drift due to changes
in ambient temperature. The two
selections for conversion loss mode
are:
periodically compensates for detector drift due to changes in
temperature. When
is selected, the detectors are automatically
zeroed approximately every five minutes. A pop-up message will briefly
appear on the analyzer’s display when the detectors are zeroed.
In manual zero mode, the analyzer zeros the detectors only when the
Zero
is pressed.
N O T E
When zeroing the detectors, the analyzer turns off the internal RF source. If you are measuring an
external source with the broadband detectors, use
the external source is
connected to the TRANSMISSION port.
6-15
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...