Using Instrument Functions
Using limit Testing
Example 2.
In this example, the analyzer has been set up with the following parameters:
Start frequency = 90 MHz
Stop frequency = 210 MHz
Number of points = 11
Maximum limit line begin frequency = 90 MHz
Maximum limit line end frequency = 200 MHz
Refer to the illustration below for the discussion. Here, the data trace passes
through the limit line, and is above it, but this test passes because the last
point
is at 198 MHz. The last point (210 MHz) does not cause the limit
test to fail because it is past the limit line stop frequency of 200 MHz. But
there is an area on the data trace between 198 MHz and 200 MHz that is
above the displayed limit line.
E N D L I M I T
a t 2 0 0 M H z
M A X I M U M
D A T A
L I M I T L I N E
I N Q U E S T I O N
N E X T
D A T A P O I N T
a t 2 1 0 M H z
l ’
I
A
9 0
1 0 2 1 1 4 1 2 6 1 3 8 1 5 0 1 6 2 1 7 4 1 8 6 1 9 8
D A T A P O I N T S
Figure 4-18. limit lines Example 2
4-46
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...