Using Instrument Functions
Using Reference Tracking
1. If you want to move the reference position (indicated by the symbol on
the left side of the display), press
Position and
then use the front panel knob, the
keys, or the numeric keypad to
enter a new reference position. A number is assigned to each possible
reference position. See Figure 4-19.
Press
Ref
Set
Frequency Use the
front panel knob, the
keys, or the numeric keypad to enter the
frequency of interest.
Press
Reference
. The reference
level is now adjusted with each sweep so that the frequency of interest
falls on the reference line. Note that the y-axis is now a relative scale,
with all amplitude values referenced to the reference line. Also note that
a small arrow appears underneath the measurement trace at the point
where the frequency of interest intersects the reference line.
4. To turn reference tracking off, press
Ref
Tracking Off .
The y-axis reverts to its previous scale format, and the reference level is
readjusted so that the measurement trace is placed correctly on the display
for absolute measurements.
4 - 4 9
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...