Using Instrument Functions
Using Markers
Using the Next Peak and Next Min Functions.
As explained previously, pressing
and
Min will
place a marker on the maximum and minimum points on the measurement
trace, respectively. You can search for the next highest or lowest point
using the
Peak Right ,
,
Min
and
keys.
A maximum (or minimum) point is detected whenever an amplitude
excursion greater than half of a division occurs. The half of a division
excursion requirement must be satisfied on both sides (left and right) of the
peak (or minimum). The maximum (or minimum) point must be
See Figure 4-4.
X Does Not Meet
Peak Search
Criteria
Figure 4.4. Peak and Minimum Search Criteria
Division
When the maximum or minimum point is at or near either edge of the
display, the excursion requirement is satisfied by a half of a division excursion
on just one side of the maximum (or minimum). See Figure 4-5.
4 - 9
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...