I
I
Using Instrument Functions
Using limit Testing
Flatness
1. This limit test requires that you
a segment on the measurement
trace using markers 1 and 2 (or markers 3 and 4 for measurement channel
2). Then press
Marker
Math
to enable the statistics marker search. See
Use Marker Flatness”
earlier in this chapter for more information on using marker statistics
2. Press
Limit Menu
Limits .
3. Use the front panel knob or the
keys to select Flatness in the
marker limit test table. Turn this limit function on by pressing the
Limit
OFF key. Note that the entry in the on/off column of the
table changes to
4. Press Edit Limit
Max Limit and enter the maximum
limit using the front panel keypad and terminating the entry with the
key.
Press Edit Limit
Limit and enter the minimum limit
using the front panel keypad and terminating the entry with the
N O T E
Note that there are no visible limit lines or indicators on the display with this limit test function.
Therefore the
Limit
OFF
function has no effect when using marker limits.
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...