I
I
Using Instrument Functions
Using limit Testing
Statistical Mean
1. This limit test requires that you first
a segment on the measurement
trace using markers 1 and 2 (or markers 3 and 4 for measurement
channel 2). Then press [MARKER)
Math
to enable the statistics marker search. See
Use Marker
Statistics” earlier in this chapter for more information on using marker
statistics.
2. Press
Limit
.
3.
the front panel knob or the
keys to select Statistic:
in the marker limit test table. Turn this limit function on by pressing the
OFF key. Note that the entry in the on/off column of the
table changes to “on.”
4. Press Edit Limit
Max Limit and enter the maximum
limit using the front panel keypad and terminating the entry with the
k
e
y
.
Press
Limit
Min Limit and enter the minimum limit
using the front panel keypad and terminating the entry with the
key.
N O T E
Note that there are no visible limit lines or indicators on the display with this limit test function.
Therefore the
Line
function has no effect when using marker limits.
4-36
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...