I
This section shows you how to determine if your measurement system
requires a
calibration.
l
Your test doesn’t require the best accuracy possible.
l
Your test device is connected directly to the reflection port with no adapters
or intervening cables.
l
Your test device impedance matches the impedance of the analyzer.
If your test setup meets these conditions, you do not need to perform any
additional calibrations, however without a user-calibration, the analyzer is not
guaranteed to meet its published measurement port speciiications.
l
You want the best accuracy possible.
l
You are adapting to a different connector type or impedance.
l
You are connecting a cable between the test device and the analyzer test
ports.
l
You are measuring a narrowband or electrically long device.
l
You are connecting any attenuator or other such device on the input or
output of the test device.
6-7
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Summary of Contents for 8711C
Page 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Page 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Page 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...