8.1 Ext. I/O Connectors and Signals
95
8
The EXT. I/O connector on the rear of the instrument supports the external control by providing sig-
nals for the end-of-measurements and comparator judgment signals, and by accepting the input of
the measurement trigger and key-lock signals.
All signals are isolated by optocouplers. (inputs and outputs share a common signal ground.)
Confirm input and output ratings, understand the safety precautions for connecting a control sys-
tem, and use accordingly.
External Control
Chapter 8
Signal input/output
Connect the EXT. I/O connector of
this instrument with the signal
output or input device
Setting the Instrument
(p. 105)
8.1 Ext. I/O Connectors and Signals
To avoid electric shock or damage to the equipment, always observe the following
precautions when connecting to the EXT. I/O connector.
•
Always turn off the power to the instrument and to any devices to be connected
before making connections.
•
During operation, a disconnected wire coming in contact with another conductive
object may cause a serious hazard. Make sure that the connections are secure and
use screws to secure the external connectors.
•
Properly insulate any devices and instruments to be connected to the EXT. I/O con-
nector.
To avoid damage to the instrument, observe the following cautions:
• Do not apply voltage or current to the EXT. I/O connectors that exceeds their ratings.
• When driving relays, be sure to install diodes to absorb counter-electromotive force.
• Be careful not to short-circuit ISO_5V to ISO_COM.
See:
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......