![Hioki RM3542-50 Instruction Manual Download Page 102](http://html1.mh-extra.com/html/hioki/rm3542-50/rm3542-50_instruction-manual_2137361102.webp)
8.1 Ext. I/O Connectors and Signals
96
• The 0ADJ signal should be asserted (Low) for at least 10 ms.
• The connector shell is conductively connected to the metal instrument chassis and the protective earth pin
of the power plug. Be aware that it is not isolated from ground.
Connector Type and Signal Pinouts
EXT. I/O Connector (Instrument Side)
Connector: (Instrument Side)
•
37-pin D-sub female with #4-40 inch
screws
Mating Connectors:
•
DC-37P-ULR (solder type)
•
DCSP-JB37PR (pressure weld type)
Japan Aviation Electronics Industry Ltd.
TRIG
(R
eserved)
K
EY_LOCK
(R
eserved)
(R
eserved)
0AD
J
HOLD
IS
O_5V
IS
O_COM
ERR
HI
LO
CE_H
I
(R
eserved)
(R
eserved)
(R
eserved)
(R
eserved)
(R
eserved)
(R
eserved)
(Re
ser
ved)
CAL
(Re
ser
ved)
(Re
ser
ved)
PRB_CHECK
(Re
ser
ved)
ISO_
COM
EO
M
INDEX
IN
PRB_SHORT
CE_LO
(Re
ser
ved)
(Re
ser
ved)
(Re
ser
ved)
(Re
ser
ved)
ISO_
12V
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
Pin
Signal name
I/O
Functions
Logic
Pin
Signal name
I/O
Functions
Logic
1
TRIG
IN
External trigger
Pos/
Neg
Edge
20 (Reserved)
−
−
−
−
2
(Reserved)
−
−
−
−
21 CAL
IN
Execute self-
calibration
Neg Edge
3
KEY_LOCK
IN
Key-Lock
Neg
Level
22 (Reserved)
−
−
−
−
4
(Reserved)
−
−
−
−
23 (Reserved)
−
−
−
−
5
(Reserved)
−
−
−
−
24 PRB_CHECK
IN
Probe short-circuit
detection Execute
Neg Edge
6
0ADJ
IN
Execute
zero-adjustment
Neg
Edge
25 (Reserved)
−
−
−
−
7
HOLD
IN
Enable external
triggering
Neg
Level
26 PRINT
IN
Print measurement
value
Neg Edge
8
ISO_5V
-
Isolated 5 V power
output
−
−
27 ISO_COM
-
Isolated common
signal ground
−
−
9
ISO_COM
-
Isolated common
signal ground
−
−
28 EOM
OUT
End of measurement
Neg Edge
10 ERR
OUT
Measurement faults Neg
Level
29 INDEX
OUT
Analog measurement
finished
Neg Edge
11 HI
OUT
Comparator
judgment result HI
Neg
Level
30 IN
OUT
Comparator
judgment result IN
Neg Level
12 LO
OUT
Comparator
judgment result Lo
Neg
Level
31 PRB_SHORT OUT
Probe short-circuit
error
Neg Level
13 CE_HI
OUT
Probe (HI side)
contact error
Neg
Level
32 CE_LO
OUT
Probe (LO side)
contact error
Neg Level
14 (Reserved)
−
−
−
−
33 (Reserved)
−
−
−
−
15 (Reserved)
−
−
−
−
34 (Reserved)
−
−
−
−
16 (Reserved)
−
−
−
−
35 (Reserved)
−
−
−
−
17 (Reserved)
−
−
−
−
36 (Reserved)
−
−
−
−
18 (Reserved)
−
−
−
−
37 ISO_12V
−
Isolated 12 V power
output
−
−
19 (Reserved)
−
−
−
−
Reserved pins are not connected inside the instrument.
Do not connect to the reserved pins.
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......