Chapter 10 Specifications
184
(7) Faulty Measurement Detection
Out-of-Range Detection Function
Operation
Indicates under- or over-range values in the following conditions:
•
Measurement value is outside of the measurement range
•
The relative calculation value is outside of the display range
•
Measurement value is outside of the A/D converter input range
•
The zero-adjusted value is outside of the display range
Ex.: Zero-adjustment in the 1
Ω
range with 0.5
Ω
zero-adjustment in effect, measur-
ing 0.1
Ω
gives a calculation result of -0.4
Ω
, which is outside of the display range.
Contact Check Function
Operation
Checks the connections between H
POT
and H
CUR
, and between L
POT
and L
CUR
termi-
nals (for each range)
Default setting
ON, 200
Ω
Setting
ON/ OFF
Threshold setting
50
Ω
/ 100
Ω
/ 150
Ω
/ 200
Ω
/ 300
Ω
/ 400
Ω
/ 500
Ω
(reference value)
Current Monitor Function
Operation
Detects faults that can obstruct normal measurement current. This function cannot be
disabled.
Voltage Level Monitor Function
Operation
Detects an error when the detection voltage is unstable (for each range)
Default setting
ON, NORMAL (LOOSE is the default only for the 100 M
Ω
range)
Setting
ON/ OFF/ ALLOFF
Threshold setting
LOOSE
NORMAL
SEVERE
Comparator Judgment and Display Indications
Table 3. Measurement States
and Display Indications
Current monitor
PASS
FAIL
Contact check
voltage level monitor
PASS
Measurement value is
displayed
Normal judgment
Out-of-range display
Hi/Lo judgment
FAIL
Contact Error
No judgment
Contact Error
No judgment
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......