4.2 Setting Pre-Measurement Delay
45
4
Estimated Delay time
Set the delay so that inductance does not affect measurements.
To fine tune the delay, begin with a longer delay than necessary, then gradually shorten it while watching the
measurement value.
1
Open the Basic Settings screen.
2
Open the Measurement Settings Screen.
3
Set [DELAY1] or [DELAY2].
4
Return to the Measurement screen.
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
1
Selection
2
Selection
1
2
Setting range: 0.0 ms (default) to 100.0 ms
[DELAY1] is common to all ranges, while
[DELAY2] can be set for each range inde-
pendently (p. 43).
Numerci keypad
3
Returns to the setting screen.
Saves and return to previous screen.
Discards setting and return to previ-
ous screen.
The confirmation screen appears.