8.6 Supplied Connector Assembly
108
The EXT. I/O connector and cover are supplied with the instrument. Assemble as shown below.
• Use shielded cables to connect a PLC to the EXT. I/O connector.
Using non-shielded conductors may result in system errors from electrical noise.
• Connect the shield to the ISO_COM pin of the EXT. I/O connector.
8.6 Supplied Connector Assembly
E
Nut #4-40UNC
Required tools:
•
Screwdriver
•
Shielded cable
•
Soldering iron
Accessories
•
A
Cover ...........................................2
•
B
Screws (+/-) #4-40UNC
(16.9 mm overall length) ..............2
•
C
Screws (+/-) #4-40UNC
(12.6 mm overall length) ..............2
•
D
Screws (
−
) #4-40UNC
(15.0 mm overall length) ..............2
•
E
Nuts #4-40UNC ............................2
•
F
Cable Clamps (for cable) .............2
•
G
Cable Clamps (case protection) ...2
•
H
Connector ....................................1
B
Screws (+/
−
) #4-40UNC (16.9 mm overall length)
A
Cover
(Identical halves)
Shielded Cables
(recommended)
C
Screw (+/
−
) #4-40UNC
(12.6 mm overall length)
D
Screw (
−
) #4-40UNC
(15.0 mm overall length)
B
Screw (+/
−
)
#4-40UNC
(16.9 mm overall length)
E
Nut #4-40UNC
D
Screw (
−
) #4-40UNC
(15.0 mm overall length)
A
Cover
(Identical halves)
F
Cable Clamps
(for cable)
G
Cable Clamps
(case protection)
Assembly Sequence
1. Solder the (shielded) cable wires to the supplied EXT. I/O connector (H) pins.
2. Affix the cable clamps (F) on the cable with screws (C).
3. Position the cable clamps (F) to fit properly inside the cover (A).
4. Insert screws (D) through the saddle washers (G).
5. In one half of cover (A), place connector (H), clamps (F), saddle washers (G) and screws (D).
6. Place the other half of cover (A) on top.
7. Affix the halves of the cover (A) together with screws (B) and nuts (E).
Be careful not to overtighten the screws, which could damage the covers.
H
Connector
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......