5.3 Power Line Frequency Manual Setting
71
5
For proper electrical noise suppression, the instrument needs to be set to match the power line fre-
quency.
With the default setting (AUTO), the instrument attempts to automatically detect the line frequency,
but manual setting is also available.
Unless the line frequency is set correctly, measurement values may be unstable.
An error message appears if line noise is high enough to prevent correct frequency detection
(ERR:041 (p. 197)). In that case, set the instrument's line frequency manually.
When the [AUTO] setting is selected, the line frequency is automatically set to 50 or 60 Hz when the instru-
ment is turned on or reset.
However, automatic detection is not available when the line frequency changes after turning power on or
resetting.
If the actual line frequency deviates from 50 or 60 Hz, it is set to one of the closest frequencies.
Examples: If the actual line frequency is 50.8 Hz, select the 50 Hz setting.
If the actual line frequency is 59.3 Hz, select the 60 Hz setting.
5.3 Power Line Frequency Manual Setting
1
Open the Basic Settings screen.
2
Open the System screen.
3
Select the line frequency to be used.
4
Return to the Measurement screen.
The Basic Settings screen appears.
The System screen appears.
[SYSTEM]
1
Selection
2
Selection
1
2
Automatically detect local line frequency
(default).
When the line frequency is 50 Hz.
When the line frequency is 60 Hz.
Returns to the setting screen.
Saves setting and return to previous screen.
Discards setting and return to previ-
ous screen.
The confirmation screen appears.
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......