Chapter 10 Specifications
182
(2) Range Switching
(3) Low-Power Function
REF% Mode
Reference value range
0.00 m
Ω
to 120.00 M
Ω
(LOW POWER: OFF)
0.0 m
Ω
to 1200.0
Ω
(LOW POWER: ON)
Upper/Lower limit ranges
-9.999% to +9.999% (when |Upper limit value| and |Lower limit value| are less than
10%)
-99.99% to +99.99% (when |Upper limit value| or |Lower limit value| is 10% or greater)
ABS Mode
Upper/Lower limit ranges
0.00 m
Ω
to 120.00 M
Ω
(LOW POWER: OFF)
0.0 m
Ω
to 1200.0
Ω
(LOW POWER: ON)
Comparator On
As per following table
Comparator Off
The range is selected by Range key
Default setting
100 M
Ω
range
Table 1. Ranges of reference (REF%) tolerance and upper limit (ABS) values
Manual setting is available on the Basic Settings screen.
VOLTAGE LIMIT: OFF (p. 64)
LOW POWER: OFF (p. 28)
VOLTAGE LIMIT: ON (p. 64)
LOW POWER: ON (p. 28)
Range of reference (REF%)
and
upper limit (ABS) values
Selected
range
Range of reference (REF%)
and
upper limit (ABS) values
Selected
range
Range of reference (REF%)
and
upper limit (ABS) values
Selected
range
0.00 m
Ω
to 100.09 m
Ω
100 m
Ω
0.00 m
Ω
to 100.09 m
Ω
100 m
Ω
−
−
100.1 m
Ω
to 1000.9 m
Ω
1000 m
Ω
100.1 m
Ω
to 1000.9 m
Ω
1000 m
Ω
100.1 m
Ω
to 1000.9 m
Ω
1000 m
Ω
1.001
Ω
to 3.009
Ω
3
Ω
1.001
Ω
to 3.009
Ω
3
Ω
1.001
Ω
to 3.009
Ω
3
Ω
3.010
Ω
to 10.009
Ω
10
Ω
3.010
Ω
to 10.009
Ω
10
Ω
3.010
Ω
to 10.009
Ω
10
Ω
10.01
Ω
to 100.09
Ω
100
Ω
10.01
Ω
to 100.09
Ω
100
Ω
10.01
Ω
to 100.09
Ω
100
Ω
100.1
Ω
to 300.9
Ω
300
Ω
100.1
Ω
to 300.9
Ω
300
Ω
100.1
Ω
to 300.9
Ω
300
Ω
301.0
Ω
to 1000.9
Ω
1000
Ω
301.0
Ω
to 1000.9
Ω
1000
Ω
301.0
Ω
to 1200.0
Ω
1000
Ω
1.001 k
Ω
to 10.009 k
Ω
10 k
Ω
1.001 k
Ω
to 5.009 k
Ω
10 k
Ω
−
−
10.01 k
Ω
to 30.09 k
Ω
30 k
Ω
5.01 k
Ω
to 15.09 k
Ω
30 k
Ω
−
−
30.10 k
Ω
to 100.09 k
Ω
100 k
Ω
15.10 k
Ω
to 50.09 k
Ω
100 k
Ω
−
−
100.1 k
Ω
to 300.9 k
Ω
300 k
Ω
50.1 k
Ω
to 150.9 k
Ω
300 k
Ω
−
−
301.0 k
Ω
to 1000.9 k
Ω
1000 k
Ω
151.0 k
Ω
to 500.9 k
Ω
1000 k
Ω
−
−
1.001 M
Ω
to 3.009 M
Ω
3 M
Ω
0.501 M
Ω
to 1.509 M
Ω
3 M
Ω
−
−
3.010 M
Ω
to 10.009 M
Ω
10 M
Ω
1.510 M
Ω
to 5.009 M
Ω
10 M
Ω
−
−
10.01 M
Ω
to 30.09 M
Ω
30 M
Ω
5.01 M
Ω
to 15.09 M
Ω
30 M
Ω
−
−
30.10 M
Ω
to 120.00 M
Ω
100 M
Ω
15.10 M
Ω
to 120.00 M
Ω
100 M
Ω
−
−
Operation
Changes resistance measurement current (setting affects all ranges)
Default setting
LOW POWER: OFF
Setting
ON/ OFF
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......