![Hioki RM3542-50 Instruction Manual Download Page 166](http://html1.mh-extra.com/html/hioki/rm3542-50/rm3542-50_instruction-manual_2137361166.webp)
9.7 Message Reference
160
Set and Query Integration Time (in seconds)
Syntax
Command
:RESistance:APERture
<Range>
,
<Speed>
,
< Integration time
(seconds)
>
Query
Response
:RESistance:APERture?
<Range>
,
<Speed>
<Integration time (seconds)>
Command
:LPResistance:APERture
<Range>
,
<Speed>
,
<Integration
Time (seconds)
>
Query
Response
:LPResistance:APERture?
<Range>
,
<Speed>
<Integration time (seconds)>
<Range> = (:RESistance ) RNG100MIL/ RNG1000MIL/ RNG3/ RNG10/ RNG100/
RNG300/ RNG1000/ RNG10K/ RNG30K/ RNG100K/ RNG300K/ RNG1000K/
RNG3MEG/ RNG10MEG/ RNG30MEG/ RNG100MEG
(:LPResistance ) RNG1000MIL/ RNG3/ RNG10/ RNG100/ RNG300/
RNG1000
<Speed> = SLOW/ MEDium/ FAST
<Integration time (seconds)> = 0.0001 sec to 0.1000 (NR2) sec
Description
Sets the integration time, in seconds.
Set and Query Contact Improver operating mode
Syntax
Command
:RESistance:CIMProve
<Range>
,
<OFF/ HOLD/ PULSe>
Query
Response
:RESistance:CIMProve?
<Range>
<OFF/ HOLD/ PULSE>
Command
:LPResistance:CIMProve
<Range>
,
<OFF/ HOLD/ PULSe>
Query
:LPResistance:CIMProve?
<Range>
Response
<OFF/ HOLD/ PULSE>
<Range> = (:RESistance ) RNG100MIL/ RNG1000MIL/ RNG3/ RNG10/ RNG100/
RNG300/ RNG1000/ RNG10K/ RNG30K/ RNG100K/ RNG300K/ RNG1000K/
RNG3MEG/ RNG10MEG/ RNG30MEG/ RNG100MEG
(:LPResistance ) RNG1000MIL/ RNG3/ RNG10/ RNG100/ RNG300/
RNG1000
Description
Sets Contact Improver operation to OFF/ HOLD/ PULSE.
Set and Query Contact Improver level
Syntax
Command
:RESistance:CIMProve:LEVel
<Range>
,
<Level>
Query
Response
:RESistance:CIMProve:LEVel?
<Range>
<Level>
Command
:LPResistance:CIMProve:LEVel
<Range>
,
<Level>
Query
Response
:LPResistance:CIMProve:LEVel?
<Range>
<Level>
<Range> = (:RESistance ) RNG100MIL/ RNG1000MIL/ RNG3/ RNG10/ RNG100/
RNG300/ RNG1000/ RNG10K/ RNG30K/ RNG100K/ RNG300K/ RNG1000K/
RNG3MEG/ RNG10MEG/ RNG30MEG/ RNG100MEG
(:LPResistance ) RNG1000MIL/ RNG3/ RNG10/ RNG100/ RNG300/
RNG1000
<Level> =
L1/ L2/ L3/ L4 (Applied current)
(L1: 17 mA/ L2: 25 mA/ L3: 35 mA/ L4: 50 mA)
Description
Sets the level of current applied by the Contact Improver function.
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...