Appendix 9 Zero Adjustment
A
18
To perform zero adjustment using a probe
When you actually perform zero adjustment using a probe, you may unexpectedly create the
connection shown in Table 1 (b). Therefore, when performing zero adjustment, you need to
pay sufficient attention to the connection state of each terminal.
9287-10 Clip Type Lead Here, is used as an example for the explaining the connection.
Table 2 shows the connection state of the tip of the lead and equivalent circuit in the respec-
tive correct and wrong connections. As indicated above, a correct connection method will be
as shown in Table 1 (a), and the voltage between H
POT
and L
POT
will be 0 V. However, a
wrong connection will be as shown in Table 1 (b), and the voltage between H
POT
and L
POT
will not be 0 V.
Table 2: Clip type lead connection methods used during zero adjustment
Connection Methods
Correct
Wrong
Tip of lead
Equivalent circuit
Deformed equivalent
circuit
As connection
method for zero
adjustment
Correct
Wrong
Red
Black
Red
Black
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......