9.7 Message Reference
140
(2) Measurement-Related
Set and Query the resistance measurement method
Syntax
Command
[:SENSe:]FUNCtion
<RESistance/ LPResistance>
Query
Response
[:SENSe:]FUNCtion?
<RESISTANCE/ LPRESISTANCE>
<RESISTANCE> = Normal Resistance measurement
<LPRESISTANCE> = Low-Power Resistance measurement
Example
FUNC LPR
Selects the Low-Power Resistance measurement.
FUNC?
RESISTANCE
The Resistance measurement has been selected.
Note
[:SENSe:] may be omitted.
Set and Query the Range Setting
Low-Power Resistance Measurement Range
Syntax
Command
[:SENSe:]LPResistance:RANGe
<Expected measurement
value>
Query
Response
[:SENSe:]LPResistance:RANGe?
<Measurement ranges (NR3)>
<Expected measurement value> = 0 to 1200
<Measurement ranges (NR3)> = 1000.000E-3/ 3.0/ 10.0/ 100.0000E+0/
300.000E+0/ 1000.000E+0
Description
Command
Enter the expected measurement value. The instrument is set to the most
suitable range for measuring the given numerical value data.
Query
Queries the measurement range setting.
Example
LPR:RANG?
10.0000E+0
The Low-Power Resistance measurement has been set to the 10
Ω
range.
Resistance Measurement Range
Syntax
Command
[:SENSe:]Resistance:RANGe
<Expected measurement
value>
Query
Response
[:SENSe:]RESistance:RANGe?
<Measurement ranges (NR3)>
<Expected measurement value> = 0 to 120E+6
<Measurement ranges (NR3)> = 100.0000E-3/ 1000.000E-3/ 3.0/ 10.0/
100.0000E+0/ 300.000E+0/ 1000.000E+0/
10.3/ 30.3/ 100.0000E+3/
300.000E+3/ 1000.000E+3/ 3.6/ 10.6/
30.0000E+6/ 100.0000E+6
Description
Command
Enter the expected measurement value. The instrument is set to the most
suitable range for measuring the given numerical value data.
Query
Queries the measurement range setting.
Example
RES:RANG 95
Sets the Resistance measurement to the 100
Ω
range.
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......