![Hioki RM3542-50 Instruction Manual Download Page 221](http://html1.mh-extra.com/html/hioki/rm3542-50/rm3542-50_instruction-manual_2137361221.webp)
Appendix 8 Table of Comparison Commands ADEX AX-162D / for this instrument
A
15
Appendix
Example 1
Measurement function: R mode
Measurement ranges: 10
Ω
Upper limit: 1.900
Ω
Lower limit: 1.100
Ω
Measurement speed: SLOW
Trigger after the above settings
Example 2
Measurement function: % mode
Reference resistance: 100 k
Ω
(Measurement range: 100 k
Ω
)
Upper limit: 10.00%
Lower limit: -10.00%
Measurement speed: FAST
Trigger after the above settings
AX-162D (Original mode)
RM3542A
M1
:CALCulate:LIMit:MODE ABS
RNG2
:RESistance:RANGe 10
LH1900E-3
:CALCulate:LIMit:ABS 1900E-3,1100E-3
LL1100E-3
S0
:SPEEd SLOW
E
*TRG
or
M1RNG2LH1900E-3LL1100E-3S0E
Note: Although no delimiters are shown, they must be included at the end of each command line.
AX-162D (Original mode)
RM3542A
M0
:CALCulate:LIMit:MODE REF
:RESistance:RANGe 100E+3
3
:CALCulate:LIMit:REFerence 100E+3
LH+1000E-2
:CALCulate:LIMit:P1000E-2,-1000E-2
LL-1000E-2
S1
:SPEEd FAST
E
*TRG
or
M3LH+1000E-2LL-1000E-2S1E
Note:
Although no delimiters are shown, they must be included at the end of each command line.
Even when the reference standard resistance is set on this instrument, the range is not automatically
selected.
Use the range setting command to set the same value as the reference standard resistance.
When setting the same value as the reference standard resistance using the range setting command, there
are cases in which the range is different from the instrument's auto-ranging selection, depending on the
value.
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......