Chapter 10 Specifications
187
10
(21) Reset Functions
(22) Self-Calibration Function
(23) Memory Function
(24) Auto-Memory Function
(25) Statistical Calculation Function
Reset
Operation
Resets settings (except the clock) to factory defaults
∗
RST (Remote command)
Operation
Resets settings (except the clock and interface) to factory defaults
When power is restored, reverts to the settings before
∗
RST was sent
Operation
Compensates for offset voltage and gain of measurement circuit
Compensation timing
When setting is changed and once every 10 minutes
Operation
Measurement values are stored by an EXT. I/O TRIG signal or by pressing the F4
[MANU]
*
button
* Displayed when EXT is selected on the [MENU]-[TRG] selection screen
Default setting
OFF
Setting
ON/ OFF
Memory capacity
30,000 (volatile memory, no backup)
Operation
Acquire once the measured value has stabilized using internal continuous trigger man-
ual measurement.
A beep will sound once the set count is reached.
Memory is cleared when acquired data has been transferred or printed by RS-232C
During Auto-Memory operation, the statistical calculation function is always enabled.
The voltage level monitor function is disabled.
Default setting
OFF/ 10
Setting
ON/ OFF
Number of measurements stored: 1 to 99
Operation
Statistical calculations are performed on measurement values in memory.
Default setting
OFF
Setting
ON/ OFF
Calculations
Total data count, Mean, Minimum value (sample no.), Maximum value (sample no.),
Standard deviation of sample, Population standard deviation, and Process capability in-
dices
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......