Chapter 10 Specifications
188
(26) Settings Monitor Function
(27) Retry Function
(28) Applied Voltage Limiter Function
(29) Self-Test Function
(30) Trigger Source Setting
(31) Measurement Speed
(32) Key-Lock Function
Operation
Measurement settings of two instruments are compared, and if different, an alarm
sounds and TRIG signal input is blocked.
However, the comparator thresholds of the 1st instrument can be smaller than those of
the 2nd.
Default setting
OFF, 1st, 0.000%
Setting
Function: ON / OFF
Measurement stages: 1st / 2nd
Tolerance range: 0.000% to 9.999%
Compared contents
Comparator threshold, measurement speed
Operation
The Retry function enables measurement to be automatically retried when a measure-
ment fault occurs due to probe chatter.
If a measurement fault persists after the specified continuous retry interval, retrying is
aborted and the EOM signal is output.
Default setting
ON, 2 ms
Setting
Function ON/ OFF
Continuous retry interval 1 ms to 50 ms
Operation
Limits the voltage applied for resistance measurement (setting affects all ranges).
Default setting
OFF
Setting
ON/ OFF
At power-on
ROM/RAM check, non-volatile ROM checksum test
Default setting
EXT (external)
Settings
INT (Internal)/ EXT (external)
Default setting
FAST
Setting
FAST/ MED/ SLOW
Operation
Disables operation of unneeded keys.
Setting
(1) Disables all except the comparator setting and cancel keys
(2) Disables all except the cancel key
(3) All front panel keys are disabled when the KEY_LOCK signal is received.
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......