![Hioki RM3542-50 Instruction Manual Download Page 168](http://html1.mh-extra.com/html/hioki/rm3542-50/rm3542-50_instruction-manual_2137361168.webp)
9.7 Message Reference
162
Set and Query contact check threshold
Syntax
Command
:RESistance:CONTactcheck:LEVel
<Range>
,
<Level>
Query
Response
:RESistance:CONTactcheck:LEVel?
<Range>
<Level>
Command
:LPResistance:CONTactcheck:LEVel
<Range>
,
<Level>
Query
Response
:LPResistance:CONTactcheck:LEVel?
<Range>
<Level>
<Range> = (:RESistance ) RNG100MIL/ RNG1000MIL/ RNG3/ RNG10/ RNG100/
RNG300/ RNG1000/ RNG10K/ RNG30K/ RNG100K/ RNG300K/ RNG1000K/
RNG3MEG/ RNG10MEG/ RNG30MEG/ RNG100MEG
(:LPResistance ) RNG1000MIL/ RNG3/ RNG10/ RNG100/ RNG300/
RNG1000
<Level>
=
L1/ L2/ L3 / L4/ L5/ L6/ L7
Contact Check Error Threshold:
(L1: 50
Ω
/ L2: 100
Ω
/ L3: 150
Ω
/ L4: 200
Ω
/ L5: 300
Ω
/ L6: 400
Ω
/ L7: 500
Ω
)
Description
Sets the error threshold value for contact checking.
Set and Query voltage level monitor
Syntax
Command
:RESistance:VMONitor
<Range>
,
<1/ 0/ ON/ OFF>
Query
Response
:RESistance:VMONitor?
<Range>
<ON/ OFF>
Command
:LPResistance:VMONitor
<Range>
,
<1/ 0/ ON/ OFF>
Query
Response
:LPResistance:VMONitor?
<Range>
<ON/ OFF>
<Range> = (:RESistance ) RNG100MIL/ RNG1000MIL/ RNG3/ RNG10/ RNG100/
RNG300/ RNG1000/ RNG10K/ RNG30K/ RNG100K/ RNG300K/ RNG1000K/
RNG3MEG/ RNG10MEG/ RNG30MEG/ RNG100MEG
(:LPResistance ) RNG100MIL/ RNG3/ RNG10/ RNG100/ RNG300/ RNG1000
Description
Enables/disables the voltage level monitor function.
Set and Query voltage level monitor level
Syntax
Command
:RESistance:VMONitor:LEVel
<Range>
,
<Level>
Query
Response
:RESistance:VMONitor:LEVel?
<Range>
<Level>
Command
:LPResistance:VMONitor:LEVel
<Range>
,
<Level>
Query
Response
:LPResistance:VMONitor:LEVel?
<Range>
<Level>
<Range> = (:RESistance ) RNG100MIL/ RNG1000MIL/ RNG3/ RNG10/ RNG100/
RNG300/ RNG1000/ RNG10K/ RNG30K/ RNG100K/ RNG300K/ RNG1000K/
RNG3MEG/ RNG10MEG/ RNG30MEG/ RNG100MEG
(:LPResistance ) RNG1000MIL/ RNG3/ RNG10/ RNG100/ RNG300/
RNG1000
<Level> =
L1/ L2/ L3
(L1: LOOSE Loose/ L2: NORMAL Normal/ L3: SEVERE Severe)
Description
Sets the error level for the voltage level monitor function.
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......