Chapter 10 Specifications
183
10
(4) Delay Setting
(5) OVC (Offset Voltage Compensation)
(6) Integration Time Setting
When OVC is set to ON, two measurements are made using the above integration time.
DELAY1
Operation
Adjusts the mechanical delay of trigger input and probing (setting affects all ranges)
Default setting
0.0 ms
Setting range
0.0 ms to 100.0 ms
DELAY2
Operation
Adjusts measuring object response (for each range)
Default setting
0.0 ms
Setting range
0.0 ms to 100.0 ms
Operation
Reverses measurement current polarity to eliminate offset voltage effects
Applicable ranges
LOW POWER OFF: 100 m
Ω
range to 10
Ω
range
LOW POWER ON: All ranges
Operation
Sets the detection voltage acquisition time span (for each range)
Default setting
As per following table
Setting range
0.1 ms to 100.0 ms, PLC
*
setting
1 to 5PLC: 50 Hz, 1 to 6PLC: 60 Hz
*. PLC: Power Line Cycle, One power line cycle (at 50 or 60 Hz)
Table 2. Integration Time Settings
Range
LOW POWER: OFF
LOW POWER: ON
Integration time
OVC
Integration time
OVC
FAST
MED
SLOW
FAST
MED
SLOW
100 m
Ω
0.5 ms
5.0 ms
1 PLC
ON
−
−
−
−
1000 m
Ω
0.3 ms
2.5 ms
1 PLC
ON
0.5 ms
5.0 ms
1 PLC
ON
3
Ω
0.3 ms
2.5 ms
1 PLC
ON
0.5 ms
5.0 ms
1 PLC
ON
10
Ω
0.3 ms
2.5 ms
1 PLC
ON
0.5 ms
5.0 ms
1 PLC
ON
100
Ω
0.3 ms
3.0 ms
1 PLC
OFF
0.3 ms
2.5 ms
1 PLC
ON
300
Ω
0.3 ms
3.0 ms
1 PLC
OFF
0.3 ms
2.5 ms
1 PLC
ON
1000
Ω
0.3 ms
3.0 ms
1 PLC
OFF
0.3 ms
2.5 ms
1 PLC
ON
10 k
Ω
0.3 ms
3.0 ms
1 PLC
OFF
−
−
−
−
30 k
Ω
0.3 ms
3.0 ms
1 PLC
OFF
−
−
−
−
100 k
Ω
0.5 ms
3.0 ms
1 PLC
OFF
−
−
−
−
300 k
Ω
0.5 ms
3.0 ms
1 PLC
OFF
−
−
−
−
1000 k
Ω
1.5 ms
5.0 ms
1 PLC
OFF
−
−
−
−
3 M
Ω
1.5 ms
5.0 ms
1 PLC
OFF
−
−
−
−
10 M
Ω
2.5 ms
1 PLC
1 PLC
OFF
−
−
−
−
30 M
Ω
2.5 ms
1 PLC
1 PLC
OFF
−
−
−
−
100 M
Ω
1 PLC
2 PLC
4 PLC
OFF
−
−
−
−
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......