Appendix 3 Unstable Measurement Values
A
6
(6) The Sample Becomes Warm
The maximum applied power to a measuring object by this instrument is determined as follows.
The resistance of samples with small thermal capacity can change due to heating. In such cases, set the low-
power resistance measurement to ON.
(7) Ingress of External Noise
Measurement should be conducted as far as possible from
devices emitting electric or magnetic fields such as power
cords, fluorescent lights, solenoid valves and PC displays. If
electrical noise ingress is a problem, prepare the measure-
ment leads as shown in Fig. 9.
(8) Measuring Transformers and Motors
If noise enters an unconnected terminal of a transformer or if motor rotor moves, measurements may be
unstable due to induced voltage on the measured winding. Pay attention to the treatment of unconnected ter-
minals on transformers or to motor vibration.
Range
Measurement
current
The maximum applied power
(Measuring object resistance) x (Measurement current)
2
VOLTAGE LIMIT: OFF
VOLTAGE LIMIT: ON
100 m
Ω
100 mA
1.2 mW
1.2 mW
1000 m
Ω
100 mA
12 mW
12 mW
3
Ω
33.3 mA
4 mW
4 mW
10
Ω
10 mA
1.2 mW
1.2 mW
100
Ω
10 mA
12 mW
12 mW
300
Ω
3.33 mA
4 mW
4 mW
1000
Ω
1 mA
1.2 mW
1.2 mW
10 k
Ω
1 mA
12 mW
12 mW
30 k
Ω
333 µA
4 mW
1.7 mW
100 k
Ω
100 µA
1.2 mW
0.5 mW
300 k
Ω
33.3 µA
400 µW
170 µW
1000 k
Ω
10 µA
120 µW
50 µW
3 m
Ω
3.33 µA
40 µW
17 µW
10 m
Ω
1 µA
12 µW
5 µW
30 m
Ω
333 nA
4 µW
1.7 µW
100 m
Ω
100 nA
1.2 µW
0.5 µW
LP
1000 m
Ω
10 mA
120 µW
120 µW
LP
3
Ω
3.33 mA
40 µW
40 µW
LP
10
Ω
1 mA
12 µW
12 µW
LP
100
Ω
1 mA
120 µW
120 µW
LP
300
Ω
333 µA
40 µW
40 µW
LP
1000
Ω
100 µA
12 µW
12 µW
Figure 9. Wiring to Minimize Noise Ingress
Twisted
Metal
Cover
A
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......