3.7 Judging Measurement Values (Comparator Function)
39
3
Setting example: Upper threshold 150 m
Ω
, lower threshold 50 m
Ω
To abort the setting process, press
. Settings are abandoned and the display returns to the
previous screen.
Decide According to Upper/Lower Thresholds (ABS Mode)
1
Open the absolute value threshold setting screen.
2
Set the positive tolerance.
Pressing an inoperative key during setting sounds a low-pitch beep (when the key beeper is enabled). Error mes-
sage is not displayed.
Press the units key to accept the setting and move the cursor to the lower threshold value.
3
Set the negative tolerance in the same way.
Press the units key to accept the setting and move the cursor to the upper threshold.
4
Accept the settings and return to the Measurement screen.
Upper threshold
Lower threshold
1_
15_
150_
150.0 m
Ω
(Example: 150
Selection
To Reset Numerical Values
Deletes entered digits.
This key is enabled only when entering
numerical values.
To change the value after selecting the units, use
the cursor
keys to select the item to
change, then enter the new value with the numer-
ic keys.
1
2
1
2
Selection
(Example: 50
5_
50_
50.0 m
Ω
•
Internal calculations are performed on floating-point values, and judgments round up any fraction of the
least-significant digit.
•
Displayed values of the reference and tolerances are rounded according to the selected range. Internal
calculations use data not rounded off, so judgments are based on the entered (setting) values.
•
An error message appears if you press
ENTER
with the positive tolerance < the negative tolerance.
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......