Preventive Maintenance and Installation
67
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
Deck
Using disinfectant cleaner applied to a paper towel or soft cloth dampened with
water, clean any contamination from the deck area.
Air Sensors
Using disinfectant cleaner applied to a paper towel or soft cloth dampened with
water, clean any contamination from the air sensors on the top deck. If
disinfectant is used, follow up with water to remove disinfectant residue from
air sensors lenses.
SPM and DPM
1. Remove the white sleeve from the pressure monitor by grasping it firmly
in your hand and pulling it straight forward from the device.
2. Using disinfectant cleaner applied to a paper towel or soft cloth
dampened with water, clean any contamination from the sleeve.
3. Using disinfectant cleaner applied to a paper towel, a soft cloth
dampened with water or an alcohol wipe, clean any contamination from
the metal tube extending from the center of the pressure monitor.
4. Insert the sleeve back into the monitor (groove end first) and push it in
until it stops - HOLD it here for a second or two and then push even
harder until you feel a distinctive “bump.” Release the pressure.
5. Verify that the sleeve is perpendicular to the chassis of the machine and
is not installed at a slight angle. If the sleeve is installed at angle, it is not
installed correctly and should reinstalled.
Outside of
Cabinet
Clean any contamination from the outside of the cabinet using disinfectant
cleaner applied to a paper towel or soft cloth dampened with water.