124
Calibration and Configuration
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
Maintenance Procedures
Procedure #
1. Access the Utilities mode of operation by pressing and holding
SHAKER
while switching the power on. Hold
SHAKER
until the UTLITIES menu
appears on the display.
2. Press
MODIFY PROGRAM
to move the arrow to MAINTENANCE and
then press
START
.
The following display appears:
3. Press
START
then
MODIFY PROGRAM
to move the arrow to
PROCEDURE # and the following display appears:
4. Press
START
and the following display appears:
5. Press
START
and the following display appears:
MAINTENANCE
PRINT CALIB
—> PROCEDURE
CLOCK
EXECUTED PROCEDURES
#Executed Procedures XXXX
EXECUTED PROCEDURES
#Executed Procedures XXXX
PRESS START AGAIN TO RESET TO 0
OR PRESS STOP TO EXIT
EXECUTED PROCEDURES
#Executed Procedures 0