Calibration and Configuration
125
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
Clock and Date
Setting
1. Access the Utilities mode of operation by pressing and holding
SHAKER
while switching the power on. Hold
SHAKER
until the UTLITIES menu
appears on the display.
2. Press
MODIFY PROGRAM
to move the arrow to MAINTENANCE and
then press
START
.
The following display appears:
3. Press
START
then
MODIFY PROGRAM
to move the arrow to CLOCK
and the following display appears:
Note: Follow the convention dd/mm/yy for the date setting. Follow the military
time convention for the time setting. The time will stop during the setting
process.
4. Use
MODIFY PROGRAM
to move the cursor to select the Date or the
Time.
5. Use
YES
and
NO
to change selection.
6. Press
START
and the clock restarts.
7. Press
STOP
. The following screen is displayed:
8. Press
STOP
and the device restarts and performs the self test.
Note: To verify that calibration procedure values were stored correctly, proceed
to DIAGNOSTICS in the UTLITIES menu. Perform all the tests listed in the
diagnostics section and record all values before using the device. For more
information, see
“Diagnostic Testing” on page 74
MAINTENANCE
PRINT CALIB
PROCEDURE
—> CLOCK
MACHINE REAL TIME CLOCK
Date: XX/XX/XX
Time: XX:XX:XX
UTILITIES
CALIBRATION
DIAGNOSTICS
CONFIGURATION
—> MAINTENANCE