Principles of Operation
47
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
DPM and SPM Offset
The DPM and SPM transducer offsets are measured and compared to a range
acceptable for ambient pressure (1902–2194 digital units). If the measured
offset is greater than the acceptable range, the test will fail. One of the following
error messages will be displayed:
SPM FAILED 20% ANALOG TEST
CODE = 11
PLEASE POWER OFF.
SPM FAILED 50% ANALOG TEST
CODE = 11
PLEASE POWER OFF.
SPM FAILED 80% ANALOG TEST
CODE = 11
PLEASE POWER OFF.
DPM OUT OF TOLERANCE
CODE = 514
PLEASE POWER OFF.
SPM OUT OF TOLERANCE
CODE = 515
PLEASE POWER OFF.