110
Calibration and Configuration
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
Market Configuration Procedure – International/North
America
Configuration is used to set up programmable features.
1. Access the Utilities mode of operation by pressing and holding
SHAKER
while switching the power on. Hold
SHAKER
until the UTLITIES menu
appears on the display.
2. Press
MODIFY PROGRAM
to move the arrow to
CONFIGURATION
and
then press
START
. The device displays:
3. Switch the memory protect switch located toward the bottom of the CPU
board (the left-most card cage board, looking from the rear of the unit) to
the Protect Off mode (down). The device displays:
4. Press
MODIFY PROGRAM
until
MACHINE
is selected and press
START
. The device displays:
CONFIGURATION
—> PROTOCOLS CONFIGURATION
LANGUAGE
******************************
**
REPLACE MEMORY
**
**
PROTECTION
**
******************************
CONFIGURATION
PROTOCOLS CONFIGURATION
LANGUAGE
—> MACHINE
MACHINE CONFIGURATION
SERIAL #
RS232 DEVICE
DATA LOGGING
—> MARKET