Disassembly
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
b. Roller
c. Rotor Top
Note: If you have any binding while pushing in the dowel pins, remove the
dowel pins and inspect for damage. Replace the dowel pin or roller as needed.
8. Inspect base of rotor to ensure dowel pins are securely fitted and not
sitting proud of rotor base. The dowel pin is normally slightly pushed into
the rotor bottom.
9. Ensure rollers and guides spin freely and do not bind. If not they must be
removed and reworked per this procedure or replace the rotor.
10.Fit rotor to pump shaft, the rotor should easily slip onto the pump shaft.
11.Align the drive pin of the pump shaft with the corresponding slot on the
rotor. Attach the rotor retaining screw.
Correctly Assembled
Incorrectly Assembled
SPM/DPM
Detector
Assembly
Note: With a thin wall socket and a bit of experience, you may be able to
remove these detectors without removing the top deck. Otherwise, see
37606-00 rotor
This example has rollers
installed beside each other.