Preventive Maintenance and Installation
79
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
13.Press
NO
to continue. The screen displays:
14.Press
YES
to continue. The screen displays:
The centrifuge starts spinning at 8000 rpm. Once the centrifuge has
stopped the following display will appear:
The device continues beeping.
BOWL SEATING TEST
IS CENTRIFUGE COVER LOCKED
PRESS YES TO CONTINUE
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240
BOWL SEATING TEST
TESTING CENTRIFUGE
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240
BOWL SEATING TEST
DID BOWL PRODUCE UNUSUAL NOISE
PRESS YES OR NO TO CONTINUE
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240