Preventive Maintenance and Installation
101
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
3. Set the meter switch settings as follows:
Leakage
Neutral closed
Polarity normal
4. Turn on the PCS2.
5. Note the reading in microamperes on the meter.
6. Depress the GND switch on the meter.
7. Change the polarity switch from normal to reverse.
8. Verify the readings comply with the current standard EN60601-1, refer to
the following table for acceptance criteria:
Ground
Continuity
1. Change the meter mode to RESISTANCE.
2. Plug the ground wire (coiled wire with clip) into the grounding socket on
the leakage meter.
3. Attach the clip to the following exposed metal locations. Verify that there
is continuity.
Rear panel hardware
Front panel hardware
Pinch valve plungers
4. Verify the highest reading is less than 0.3 Ohm.
Condition
Polarity
Ground
Limit
Normal
Normal
Normal
< 100 µ amps
Single Fault
Normal
Open
< 500 µ amps
Single Fault
Reverse
Open
< 500 µ amps
Normal
Reverse
Normal
< 100 µ amps