168
Troubleshooting
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
52
CONFIGURATION
CRC
FAILURE
Configuration CRC does
not match value stored in
NVRAM.
1. Reconfigure the language in Utilities.
2. Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
centrifuge, and configure the language.
Reset protocol parameters (
MODIFY
PROGRAM
) if parameters were
different from default values.
502
SPM FILTER
INTERLOCK
FAILURE
Fifty second time-out has
occurred after request to
remove SPM filter. Before
troubleshooting, be sure
filter was removed within
fifty seconds of displayed
request.
1. If request to remove filter remains
displayed after removing SPM filter from
SPM within fifty seconds, refer to the
system schematic and check, repair or
replace all electrical connections
associated with the SPM.
2.Use Diagnostics to check SPM
interlocks. Check that filter sleeve is
properly installed in SPM and that the
filter interlocks respond properly when
the filter sleeve is twisted clockwise and
counterclockwise. Remove filter sleeve
and reseat. Replace SPM if problem is
found. Use Utilities to re-calibrate A/D.
3. Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
centrifuge, and configure the language.
Reset protocol parameters (
MODIFY
PROGRAM
) if parameters were
different from default values.
Table 9, CPU error messages